Agilent Technologies 35670A Manual page 241

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate[1|2|3|4]:WATerfall:SLICe:SELect:POINt
Selects a waterfall slice by its display point value.
Command Syntax:
<number> ::= a real number (NRf data)
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command specifies the X-axis position for the waterfall slice by point number.
The number of points displayed along the X-axis depends upon the number of lines of resolution seet with
the [SENSe:]FREQuency:RESolution command. See table 6-8.
In correlation analysis the number of points displayed along the X-axis depends upon the resolution and
the windowing function. See table 6-8.
In histogram analysis, the number of points is determined by the number of bins (set with the
[SENS:]HIST:BINS command). The maximum number is 1024 points.
In octave analysis, the number of points is determined by the bandwidth of the filters. There are 11 points
for full octave, 33 points for 1/3 octave and 132 points for 1/12 octave.
In some cases the number of points is arbitrary. These include waterfall displays from order tracking or
from the arbitrary source.
See the CALC:WAT:SLICe:COPY command for information about saving a waterfall slice to a data
register.
CALCulate[1|2|3|4]:WATerfall:SLICe:SELect:POINt <number>|
<step>|<bound>
limits:
0:2048
OUTPUT 711;"Calc:Waterfall:Slic:Sel:Point 816"
OUTPUT 711;"CALC4:WATERFALL:SLIC:SEL:POINT 1409"
CALCulate[1|2|3|4]:WATerfall:SLICe:SELect:POINt?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
CALCulate
command/query
6-133