CALCulate
CALCulate[1|2|3|4]:WATerfall:TRACe:SELect
Selects a waterfall trace by its Z-axis value.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [S|RPM|COUNT|AVG]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command selects a waterfall trace by its Z-axis value. The Z-axis value indicates when the
measurement data was armed.
To specify a waterfall trace in trace box A, send CALC1; in trace box B, send CALC2; in trace box C,
send CALC3; and in trace box D, send CALC4. The trace specifier defaults to trace box A if the specifier
is not used.
See the CALC:WAT:TRAC:COPY command for information about copying a selected trace to a data
register.
6-136
CALCulate[1|2|3|4]:WATerfall:TRACe:SELect {<number>
[<unit>]}|<step>|<bound>
limits:
-9.9e37:9.9e37
OUTPUT 711;"calculate:wat:trace:sel 10 count"
OUTPUT 711;"CALC3:WAT:TRAC:SEL 3.5 S"
CALCulate[1|2|3|4]:WATerfall:TRACe:SELect?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+0.0
SCPI Compliance:
no
instrument-specific
command/query