[SENSe:]
[SENSe:]FREQuency:SPAN
Specifies the frequency bandwidth to be measured.
Command Syntax:
<number> ::= a real number (NRf data)
limits: 0.015625:102400.0
<unit> ::= [HZ|CPM]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
In FFT analysis instrument mode (INST:SEL FFT):
The value of FREQ:SPAN is used with either FREQ:CENT or FREQ:STAR to define the band of
frequencies. The maximum and the minimum frequency spans are listed in table 18-7. Allowable
values for the frequency span are determined by the following formula:
When you send this command, the value of the record length (SWE:TIME) is adjusted so the
following formula is true:
The frequency span limits the range of the start frequency values (FREQ:STAR) according to the
following formulas:
for one channel measurements:
(FREQ:STAR)  115 kHz - (FREQ:SPAN)/2
for two channel measurements:
(FREQ:STAR)  57.5 kHz - (FREQ:SPAN)/2
for four channel measurements:
(FREQ:STAR)  28.75 kHz - (FREQ:SPAN)/2
18-40
[SENSe:]FREQuency:SPAN {<number>[<unit>]}|<step>|<bound>
OUTPUT 711;":FREQ:SPAN 5E4"
OUTPUT 711;"sens:freq:span MIN"
[SENSe:]FREQuency:SPAN?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+1.024E+005
SCPI Compliance:
no
confirmed
n
≤ ≤
command/query