DISPlay:VIEW
Specifies what is displayed on the analyzer's screen.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command specifies the contents of the analyzer's display area.
To display measurement data, send TRAC. Use the CALC:FEED commands to specify the
measurement data to be displayed in the active trace.
To display the analyzer's current measurement configuration, send MST.
To display the contents of the default disk, send MMEM. (You select the default disk with the
MMEM:MSIS command.)
To display the synthesis table, send STAB.
To display the curve fit table, send CTAB.
To display the fault log table, send FTAB.
To display the test log table, send TTAB.
To display the memory usage table, send MEM.
To display the option configuration of the analyzer, send OPT.
To display header information for the time capture buffer, send CAPT.
To display the most recent messages (up to 5) that had been displayed on the analyzer's screen, send
MESS.
DISPlay:VIEW TRACe|MSTate|MMEMory|STABle|CTABle|FTABle|
TTABle|MEMory|OPTion|CAPTure|ISTate|MESSage
OUTPUT 711;":DISP:VIEW CAPT"
OUTPUT 711;"display:view mstate"
DISPlay:VIEW?
TRAC|MST|MMEM|STAB|CTAB|FTAB|TTAB|MEM|OPT|CAPT|MESS
Option:
not applicable
Synchronization Required:
Preset State:
TRAC
SCPI Compliance:
no
instrument-specific
DISPlay
command/query
8-17