Agilent Technologies 35670A Manual page 385

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

MMEMory:LOAD:CONTinue
Continues the load operation of time capture and waterfall files saved on multiple disks.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command loads split files which were saved with the MMEM:STOR:CONT commands.
MMEM:LOAD:CONT is valid only with time capture and waterfall files.
Use the MMEM:LOAD:TCAP or MMEM:LOAD:WAT command to begin the load operation.
When the analyzer has completed loading the first split file, filename_1, it generates a message, "Media
full; Insert next disk with 'filename_2'. Insert the disk containing filename_2. Send this command to
continuing loading the file.
You can use the MMEM:LOAD:CONT? to verify that the time capture or waterfall file has been
transferred. If the MMEM:LOAD:CONT? query returns a +1, the analyzer has not completed the
transfer. The query returns a 0 when the analyzer has transferred the entire time capture or waterfall file.
For additional information on the generation of split files, see the MMEM:STOR:CONT command.
MMEMory:LOAD:CONTinue
OUTPUT 711;"MMEM:LOAD:CONT"
OUTPUT 711;"mmem:load:cont"
MMEMory:LOAD:CONTinue?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
MMEMory
command/query
15-11

Advertisement

Table of Contents
loading