[SENSe:]
[SENSe:]TCAPture:FILE
Uploads or downloads time-capture file between the analyzer and the controller.
Command Syntax:
<DATA> ::= <DEF_BLOCK>
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command is used to upload and download time-capture files between theanalyzer and the controller.
The main use of this command is for fast storage of time-capture files on a personal computer hard disc.
See the XFER example program in appendix F.
The files sent to the controller are SDF (Standard Data Format) files. The files sent to the analyzer must
also be SDF time-capture files.
Use the SENSE:DATA command to upload and download time-capture data.
18-70
[SENSe:]TCAPture:FILE
OUTPUT 711;"SENSE:TCAP:FILE?"
[SENSe:]TCAPture:FILE?
DEF_USER
Option: not applicable
Synchronization Required: no
Preset state: not applicable
SCPI Compliance: instrument-specific
command/query