Agilent Technologies 35670A Manual page 199

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate[1|2|3|4]:MARKer:X:RELative
Specifies the marker reference X-axis position relative to the main marker.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [HZ|S|ORD|COUNT|AVG|V|VPK|RPM|EU|REV]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command specifies the marker reference X-axis position relative to the main marker's position. To
query the units for the X-axis, send CALC:MARK:X:REL? UNIT.
To specify an absolute X-axis position for the marker reference, use the CALC:MARK:REF:X command.
CALCulate[1|2|3|4]:MARKer:X:RELative {<number>[<unit>]}|
<step>|<bound>
limits:
-102400.0:102400.0
OUTPUT 711;":calculate:mark:x:relative 43013.2"
OUTPUT 711;"Calc:Marker:X:Rel -43742.4"
CALCulate[1|2|3|4]:MARKer:X:RELative?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+0.0
SCPI Compliance:
no
instrument-specific
CALCulate
command/query
6-91