[SENSe:]
[SENSe:]FREQuency:STEP[:INCRement]
Specifies the step size which is used when changing frequency parameters.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [HZ|CPM]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
In FFT analysis instrument mode (INST:SEL FFT), step size determines the change in frequency which
results when you send UP or DOWN with the FREQ:CENT or the FREQ:STAR commands.
In swept sine instrument mode (INST:SEL SINE; Option 1D2), step size determines the change in
frequency which results when you send UP or DOWN with any of the following commands:
FREQ:CENTer
FREQ:STARt
FREQ:STOP
FREQ:MANual
18-48
[SENSe:]FREQuency:STEP[:INCRement] {<number>[<unit>]}|
<step>|<bound>
limits:
0.015625:102400.0
OUTPUT 711;":FREQ:STEP 5e3"
OUTPUT 711;"sens:freq:step:incr 100"
[SENSe:]FREQuency:STEP[:INCRement]?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+2.0E+003
SCPI Compliance:
no
confirmed
command/query