[SENSe:]
In correlation analysis instrument mode (INST:SEL CORR), the following commands are valid:
AVER:TYPE RMS
AVER:TYPE RMS;TCON EXP
AVER:TYPE TIME
AVER:TYPE TIME;TCON EXP
In histogram instrument mode (INST:SEL HIST):
This command is not valid.
Octave analysis instrument mode (INST:SEL OCT):
To select linear averaging, send AVER:TYPE RMS. Old and new data records are weighted
equally to yield the arithmetic mean. Averaging is done for a specified amount of time rather than
for a number of averages. The value specified with the AVER:TIME command is used for linear
integration time.
To select exponential averaging, send AVER:TYPE RMS;TCON EXP. New data is weighted more
than old data. The AVER:TIME command specifies the time constant.
To select equal confidence averaging, send AVER:TYPE ECON;TCON EXP. The averaging time
for each band is proportional to the bandwidth product for that band. The relative confidence in the
measurement is equal across bands. The AVER:CONF command determines the specifies the
confidence level. This is an instrument-specific SCPI command.
To specify the peak hold function, send AVER:TYPE MAX. The value specified with the
AVER:TIME command is the integration time over which to hold maximum values.
In swept sine instrument mode (INST:SEL SINE):
This command is not valid. Averaging integrates a single data point at a time, as opposed to
averaging complete time records.
18-20