MMEMory
MMEMory:LOAD:CFIT
Loads a curve fit table into the analyzer from a file on the specified disk.
Command Syntax:
<FILENAME> ::= '[<PATH>]<filename>'
<PATH> ::= <MSIS>[{DOS_DIR}]
<MSIS> ::= RAM:|NVRAM:|INT:|
<DOS_DIR> ::= \<DIR_NAME>\
<DIR_NAME> ::= ASCII characters
<filename> ::= ASCII characters (see beginning of chapter
Example Statements:
Attribute Summary:
Description:
This command loads a curve fit table into the curve fit buffer. The file must have been saved with the
MMEM:STOR:CFIT or MMEM:STOR:SYNT commands. We recommend saving curve fit files with the
".FIT" file extension and synthesis files with the ".SYN" file extension.
If you are loading a synthesis table into the curve fit buffer, the table must be in pole-zero format. See the
CALC:SYNT:TTYP command for information about table formats.
The current curve fit table is overwritten.
Note
The file extensions ".FIT" and ".SYN" are naming conventions. The file type is created
with the MMEM:STOR:CFIT or MMEM:STOR:SYNT commands. This file-type data
is embedded within the file.
15-10
MMEMory:LOAD:CFIT <FILENAME>
EXT[,<select_code>[,<unit _number>]]:
(DOS only)
for <filename> restrictions)
OUTPUT 711;"MMEMORY:LOAD:CFIT 'CFFILE.FIT'"
OUTPUT 711;"mmem:load:cfit 'int:synth1.syn'"
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
command