Calculate[1|2|3|4]:X:data?Query - Agilent Technologies 35670A Manual

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate[1|2|3|4]:LIMit:STATe
Turns limit testing on and off for the specified trace.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
When limit testing is on, the specified trace is evaluated against the limits defined in its upper and lower
limit registers. If a trace specifier is not used, the command defaults to trace A. You can load these
registers via the GPIB using the CALC:LIM:LOW:SEGM and the CALC:LIM:UPP:SEGM commands or
the CALC:LIM:UPP:TRAC and CALC:LIM:LOW:TRAC commands.
To determine whether or not a trace is within the specified limits, you can send the CALC:LIM:FAIL
query or monitor the bits in the Limit Fail condition register. (For more information, see "Limit Fail
Register Set" in chapter 1.)
To return failed points on the X-axis, send the CALC:LIM:LOW:REP and the CALC:LIM:UPP:REP
queries. To return failed points on the Y-axis, send the CALC:LIM:LOW:REP:YDAT and the
CALC:LIM:UPP:REP:YDAT queries.
Note
Limit lines are not automatically displayed when limit testing is enabled. To display
limits you must send DISPlay[:WINDow[1|2|3|4]]:LIMt:STATe ON.
CALCulate[1|2|3|4]:LIMit:STATe OFF|0|ON|1
OUTPUT 711;":CALC2:LIM:STATE OFF"
OUTPUT 711;"calc:limit:stat ON"
CALCulate[1|2|3|4]:LIMit:STATe?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
OFF (+0)
SCPI Compliance:
no
confirmed
CALCulate
command/query
6-43

Advertisement

Table of Contents
loading

Table of Contents