Calculate[1|2|3|4]:Marker:x[:Absolute]Command/Query - Agilent Technologies 35670A Manual

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

ARM:RPM:THReshold
Specifies the starting RPM value.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [RPM]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command is only valid when the Start RPM Arming qualifier is enabled with the ARM:RPM:MODE
command.
The starting value is either the lowest RPM used for devices with increasing RPM (ARM:RPM:MODE
UP) or the highest RPM for devices with decreasing RPM (ARM:RMP:MODE DOWN).
ARM:RPM:THReshold {<number>[<unit>]}|<step>|<bound>
limits:
5:491520
OUTPUT 711;"ARM:RPM:THR 1000"
OUTPUT 711;"arm:rpm:threshold 8000"
ARM:RPM:THReshold?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+600 RPM
SCPI Compliance:
no
instrument-specific
ARM
command/query
5-5