[SENSe:]REJect:STATe
Turns overload rejection on or off.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
If overload rejection is off (the default condition), all time records are included in the measurement.
If SENS:REJ ON is sent, the time record from the overloaded input channel is not included in the
measurement results. The concurrent time record from the other channel is rejected as well. The
measurement continues until the analyzer has collected the specified number of non-overloaded time
records.
[SENSe:]REJect:STATe OFF|0|ON|1
OUTPUT 711;":rej:stat ON"
OUTPUT 711;"Sense:Rej:State ON"
[SENSe:]REJect:STATe?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
OFF (+0)
SCPI Compliance:
no
instrument-specific