SOURce:VOLTage[:LEVel]:AUTO
Enables the autolevel feature in swept sine measurements (INST:SEL SINE).$Isource;autolevel
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
If autolevel is enabled with the SOUR:VOLT:AUTO ON command, the analyzer adjusts the source
output level to keep the amplitude of an input channel within a specified range.
See the following commands for additional information about setting the range:
SOURce:VOLTage:LIMit[:AMPLitude]
SOURce:VOLTage:LIMit:INPut
SOURce:VOLTage[:LEVel]:REFerence
SOURce:VOLTage[:LEVel]:REFerence:CHANnel
SOURce:VOLTage[:LEVel]:REFerence:TOLerance
To turn off the autolevel function, send SOUR:VOLT:AUTO OFF.
Refer to Online Help for detailed information about the autolevel feature.
SOURce:VOLTage[:LEVel]:AUTO OFF|0|ON|1
OUTPUT 711;":sour:voltage:auto OFF"
OUTPUT 711;"Source:Volt:Lev:Auto ON"
SOURce:VOLTage[:LEVel]:AUTO?
Integer
Option:
1D2 Swept Sine
Synchronization Required:
Preset State:
OFF (+0)
SCPI Compliance:
no
instrument-specific
SOURce
command/query
19-9