Agilent Technologies 35670A Manual page 158

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate
CALCulate[1|2|3|4]:LIMit:UPPer:SEGMent:CLEar
Deletes a segment from the upper limit line.
Command Syntax:
<number> ::= a real number (NRf data)
<bound> ::= MAX|MIN
Example Statements:
Attribute Summary:
Description:
This command deletes any segment which contains the X-axis value you specify. Adjacent segments are
not affected, although the limit line may be discontinuous.
The value entered for a limit line segment is unitless.
To delete all segments of an upper limit line, send CALC:LIM:UPP:CLE.
6-50
CALCulate[1|2|3|4]:LIMit:UPPer:SEGMent:CLEar <number>|
<bound>
limits: -9.9e37:9.9e37
(depend upon current range of the X-axis)
OUTPUT 711;"calculate4:lim:upp:segment:cle -3.21429e+37"
OUTPUT 711;"Calculate4:Lim:Upp:Segment:Cle -2.22796e+37"
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific