CALCulate
CALCulate[1|2|3|4]:MARKer:MAXimum:RIGHt
Moves the main marker one peak to the right of its current location on the specified trace.
Command Syntax:
Example Statements:
Attribute Summary:
Description:
The trace specifier, CALCulate[1|2|3|4], determines which trace you are selecting. Omit the specifier or
send 1 for trace A, 2 for trace B, 3 for trace C, or 4 for trace D.
A peak is a local maximum on the displayed trace. The slope of a trace is positive to the left of a peak
and negative to the right. In addition, the slope on one side of a peak must not change for at least one
vertical division (one-third division on both sides if in octave analysis instrument mode or one-tenth
division (one decade) if using logarithmic scaling of the Y-axis).
This command only finds peaks that are at least one point to the right of the current marker position. If
the analyzer does not find a peak, the marker does not move. You can increase the number of peaks
found by the analyzer by decreasing the value of vertical scale division (DISP:TRAC:Y:PDIV).
6-78
CALCulate[1|2|3|4]:MARKer:MAXimum:RIGHt
OUTPUT 711;"Calculate4:Mark:Max:Right"
OUTPUT 711;"CALC4:MARKER:MAX:RIGH"
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no