Agilent Technologies 35670A Manual page 505

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSe:]SWEep:TIME
Specifies the length of the time record in seconds.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [S]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
When you send this command, two other values may be adjusted. The value of the frequency span
(FREQ:SPAN) is adjusted to FREQ:RES / record length (in hertz). If the start frequency is "fixed"
(FREQ:SPAN:LINK STAR), the center and stop frequencies are adjusted accordingly. If the center
frequency is "fixed" (FREQ:SPAN:LINK CENT), the start and stop frequencies are adjusted.
If you change the frequency span, the time record length is adjusted to:
In correlation analysis instrument mode (INST:SEL CORR):
This command specifies the record length, T.
collected. T is used in correlation windowing functions as follows:
The record length, T, is independent of the frequency resolution (FREQ:RES). Changing the value
of one does not change the value of the other. See the online help for additional information about
record length in correlation analysis.
[SENSe:]SWEep:TIME {<number>[<unit>]}|<step>|<bound>
limits:
976.5625e-6:8192.0
OUTPUT 711;"SWE:TIME 1 S"
OUTPUT 711;"sense:sweep:time 3.125e-2"
[SENSe:]SWEep:TIME?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+7.8049E-003
SCPI Compliance:
no
confirmed
T represents the length of the raw time record
[SENSe:]
command/query
18-67