Agilent Technologies 35670A Manual page 461

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSE:]DATA:HEADer:FREQuency:STOP?
Returns the stop frequency setting used for the current time-capture buffer.
Query Syntax:
Example Statements:
Return Format:
Attribute Summary:
Description:
This command is used to determine the stop frequency used when the data in the current time-capture
buffer was captured.
The time-capture start and stop frequency can be used to calculate the time spacing between the capture
data points. The spacing is equal to 400 / ((Stop frequency) − (Start frequency)).
[SENSe:]DATA:HEADer:FREQuency:STOP?
OUTPUT 711;"SENSE:DATA:HEAD:FREQ:STOP?"
Real
Option: not applicable
Synchronization Required: no
Preset state: 0
SCPI Compliance: instrument-specific
[SENSe:]
query
18-23