INSTrument[:SELect]
Selects one of the analyzer's six major instrument modes.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The following commands select a major instrument mode:
INST:SEL CORR – Correlation Analysis
INST:SEL FFT – FFT Analysis
INST:SEL HIST – Histogram
The following commands are valid if the appropriate option is installed:
INST:SEL OCT – Octave Analysis – Option 1D1
INST:SEL ORD – Order Analysis – Option 1D0
INST:SEL SINE – Swept Sine – Option 1D2
The GPIB command set changes with each instrument mode. Parameters sent to setup a measurement for
each instrument mode do not affect setups for other instrument modes. As a result, you should select the
instrument mode near the beginning of any program sequence that defines the instrument state.
INSTrument[:SELect] FFT|OCTave|ORDer|SINE|HISTo-
gram|CORRelation
OUTPUT 711;":Inst SINE"
OUTPUT 711;"INST:SELECT FFT"
INSTrument[:SELect]?
FFT|OCT|ORD|SINE|HIST|CORR
Option:
1D0 Computed Order Tracking
1D1 Realtime Octave
1D2 Swept Sine
Synchronization Required:
Preset State:
FFT
SCPI Compliance:
no