Agilent Technologies 35670A Manual page 362

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

INSTrument
INSTrument:NSELect
Selects one of the analyzer's six major instrument modes.
Command Syntax:
<number> ::= a real number (NRf data)
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The following commands select a major instrument mode:
INST:NSEL 0 – FFT Analysis
INST:NSEL 4 – Histogram
INST:NSEL 5 – Correlation Analysis
The following commands are valid if the appropriate option is installed:
INST:NSEL 1 – Octave Analysis – Option 1D1
INST:NSEL 2 – Order Analysis – Option 1D0
INST:NSEL 3 – Swept Sine – Option 1D2
The GPIB command set changes with each instrument mode. Parameters sent to setup a measurement for
each instrument mode do not affect setups for other instrument modes. As a result, you should select the
instrument mode near the beginning of any program sequence that defines the instrument state.
The default instrument mode is FFT at powerup and reset.
13-2
INSTrument:NSELect <number>|<bound>
limits:
0:5
OUTPUT 711;"INST:NSELECT 4"
OUTPUT 711;"inst:nselect 4"
INSTrument:NSELect?
Integer
Option:
1D0 Computed Order Tracking
1D1 Realtime Octave
1D2 Swept Sine
Synchronization Required:
Preset State:
0
SCPI Compliance:
no
(FFT)
confirmed
command/query