DISPlay[:WINDow[1|2|3|4]]:TRACe:BPOWer[:STATe]
Turns on or off the display of the overall band.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command displays the overall band in octave analysis instrument mode. A letter below the band
indicates the type of power.
T indicates band-limited total power.
I indicates broadband impulse power (SENSe:AVERage:TYPE MAX).
B indicates broadband total power.
P indicates broadband peak power (SENSe:AVERage:IMPulse ON).
U indicates undefined data (data read from a different analyzer).
WINDow[1|2|3|4] specifies the trace box. Send WIND1 to specify A, WIND2 to specify B, WIND3 to
specify C, and WIND4 to specify D. If you do not send the optional trace box specifier, the command
defaults to trace box A.
In addition, you can display the overall band with the A-weight filter applied to the overall band. See the
DISPlay:WINDow:TRACe:APOWer:STATe command for more information.
DISPlay[:WINDow[1|2|3|4]]:TRACe:BPOWer[:STATe] OFF|0|ON|1
OUTPUT 711;"DISPLAY:WIND:TRACE:BPOW:STAT OFF"
OUTPUT 711;"display:trac:bpower ON"
DISPlay[:WINDow[1|2|3|4]]:TRACe:BPOWer[:STATe]?
Integer
Option:
1D1 Realtime Octave
Synchronization Required:
Preset State:
ON (+1)
SCPI Compliance:
no
instrument-specific
DISPlay
command/query
8-25