[SENSe:]
[SENSe:]SWEep:MODE
Specifies automatic or manual sweep modes.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
To select automatic sweep mode, send SWE:MODE AUTO. The instrument controls the sweep
according to the following parameters:
FREQ:STARt
FREQ:STOP
FREQ:RESolution
SWE:DIRection
SWE:SPACe
The values of the parameters are specified by default or by the appropriate command.
Send SWE:MODE MAN to select a discrete sweep; the measurement occurs only at the frequency points
specified by the [SENSe:]FREQ:MANual command.
18-62
[SENSe:]SWEep:MODE AUTO|MANual
OUTPUT 711;":Swe:Mode AUTO"
OUTPUT 711;"SENSE:SWE:MODE AUTO"
[SENSe:]SWEep:MODE?
AUTO|MAN
Option:
1D2
Swept Sine
Synchronization Required:
Preset State:
AUTO
SCPI Compliance:
no
confirmed
command/query