Agilent Technologies 35670A Manual page 281

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

DISPlay[:WINDow[1|2|3|4]]:TRACe:LABel
Loads a label for the specified trace.
Command Syntax:
<STRING> ::= ASCII characters - 32 through 126
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
Trace titles replace the default titles supplied by the analyzer. They appear above the upper-left corner of
the trace box. Trace titles can be a maximum of 13 characters long.
WINDow[1|2|3|4] specifies the trace box. Send WIND1 to specify A, WIND2 to specify B, WIND3 to
specify C, and WIND4 to specify D. If you do not send the optional trace box specifier, the command
defaults to trace box A.
Note
If you send *RST or SYST:PRES, trace titles are automatically erased. The analyzer
restores the default title for all trace boxes.
DISPlay[:WINDow[1|2|3|4]]:TRACe:LABel '<STRING>'
maximum number of characters:
OUTPUT 711;"DISPLAY:WIND3:TRACE:LABEL 'CEPSTRUM'"
OUTPUT 711;"disp:trace:lab 'SPL'"
DISPlay[:WINDow[1|2|3|4]]:TRACe:LABel?
"<STRING>"
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
13
no
instrument-specific
DISPlay
command/query
8-27

Advertisement

Table of Contents
loading

Table of Contents