Agilent Technologies 35670A Manual page 487

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

[SENSe:]FREQuency:STOP
Sets the stop frequency to the specified value.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [HZ|CPM]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
In FFT analysis instrument mode (INST:SEL FFT):
This command defines the upper limit of the frequency bandwidth. The start frequency remains
fixed (FREQ:SPAN:LINK). The values for the center frequency, the frequency span, and the record
length change to appropriate values.
In octave measurement mode (INST:SEL OCT; Option 1D1):
This command specifies the stop frequency for the octave measurement. The frequency can be
specified in Hz or as a band number.
In swept sine instrument mode (INST:SEL SINE; Option 1D2):
The value specified by this command is rounded to the next higher 15.625 mHz step. The start
frequency (FREQ:STAR) is held constant and selected as the new "anchor" for the measurement.
The center frequency and frequency span are adjusted to appropriate values.
[SENSe:]FREQuency:STOP {<number>[<unit>]}|<step>|<bound>
limits:
0.03125:115000.0
OUTPUT 711;"SENS:FREQ:STOP 25.6 KHZ"
OUTPUT 711;"frequency:stop max"
[SENSe:]FREQuency:STOP?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+1.024E+005
SCPI Compliance:
no
confirmed
[SENSe:]
command/query
18-49