STATus
STATus:QUEStionable:PTRansition
Sets and queries bits in the Questionable Status positive transition register.
Command Syntax:
<number> ::= a real number (NRf data)
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
To set a single bit in the Questionable Status positive transition register to 1, send the bit's decimal weight
with this command. To set more than one bit to 1, send the sum of the decimal weights of all the bits.
(The decimal weight of a bit is 2
All bits are initialized to 0 when the analyzer is turned on or when the STAT:PRES command is sent.
However, the current setting of bits is not modified when you send the *RST command.
See "Questionable Status Register Set" in chapter 1 for a definition of bits in the register set. See
"Programming the Status System" in the GPIB Programmer's Guide for information about the role
of positive transition registers in register sets.
20-22
STATus:QUEStionable:PTRansition <number>|<bound>
limits:
0:32767
OUTPUT 711;"STAT:QUESTIONABLE:PTR 256"
OUTPUT 711;"status:ques:ptr 768"
STATus:QUEStionable:PTRansition?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
not affected by Preset
SCPI Compliance:
n
, where n is the bit number.)