DISPlay
DISPlay[:WINDow[1|2|3|4]]:WATerfall:BOTTom
Specifies the bottom Z-axis value in a waterfall display.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [S|RPM|COUNT|AVG]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command specifies the starting Z-axis value which appears at the bottom of a waterfall display. Use
the DISP[:WIND]:WAT:TOP command to specify the corresponding Z-axis value which appears at the
top of the waterfall display.
Use the DISP[:WIND]:WAT:COUNt command to specify a range of Z-axis values. The
DISP:WAT:COUNt command determines the number of traces displayed in a waterfall. If a range of
values is specified, the analyzer holds the DISP:WAT:BOTTom value constant and adjusts the top value.
WINDow[1|2|3|4] specifies the trace box. Send WIND1 to specify A, WIND2 to specify B, WIND3 to
specify C, and WIND4 to specify D. If you do not send the optional trace-box specifier, the command
defaults to trace box A.
8-44
DISPlay[:WINDow[1|2|3|4]]:WATerfall:BOTTom {<number>
[<unit>]}|<step>|<bound>
limits:
0:9.9e37
OUTPUT 711;":DISPLAY:WAT:BOTT 3.94581e+37"
OUTPUT 711;"display:wind2:waterfall:bott 8.84371e+37"
DISPlay[:WINDow[1|2|3|4]]:WATerfall:BOTTom?
Real
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
no
instrument-specific
command/query