STATus
STATus:QUEStionable:LIMit:CONDition?
Reads and clears the Limit Fail condition register.
Query Syntax:
Example Statements:
Return Format:
Attribute Summary:
Description:
This query returns the sum of the decimal weights of all bits currently set to 1 in the Limit Fail condition
register. (The decimal weight of a bit is 2
See "Limit Fail Register Set" in chapter 1 for a definition of bits in the register set. See "Programming
the Status System" in the GPIB Programmer's Guide for information about the role of condition
registers in register sets.
20-16
STATus:QUEStionable:LIMit:CONDition?
OUTPUT 711;"STATUS:QUES:LIM:CONDITION?"
OUTPUT 711;"stat:questionable:lim:cond?"
Integer
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
n
, where n is the bit number.)
no
instrument-specific
query