Agilent Technologies 35670A Manual page 558

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

STATus
STATus:DEVice:PTRansition
Sets and queries bits in the Device State positive transition register.
Command Syntax:
<number> ::= a real number (NRf data)
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
To set a single bit in the Device State positive transition register to 1, send the bit's decimal weight with
this command. To set more than one bit to 1, send the sum of the decimal weights of all the bits. (The
decimal weight of a bit is 2
All bits are initialized to 0 when the analyzer is turned on or when the STAT:PRES command is sent.
However, the current setting of bits is not modified when you send the *RST command.
See "Device State Register Set" in chapter 1 for a definition of bits in the register set. See "Programming
the Status System" in the GPIB Programmer's Guide for information about the role of positive
transition registers in register sets.
20-6
STATus:DEVice:PTRansition <number>|<bound>
limits:
0:32767
OUTPUT 711;"status:dev:ptransition 17"
OUTPUT 711;"stat:dev:ptr 20"
STATus:DEVice:PTRansition?
Integer
Option:
not applicable
Synchronization Required:
Preset State:
not affected by Preset
SCPI Compliance:
n
, where n is the bit number.)
no
instrument-specific
command/query

Advertisement

Table of Contents
loading

Table of Contents