Agilent Technologies 35670A Manual page 638

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

TRIGger
TRIGger:LEVel
Specifies the level of the input signal which causes the analyzer to trigger.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [PCT|V|VPK|EU]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
You can specify the trigger level as a value, as a percentage of the trigger channel's current input range,
or as a fraction of the trigger channel's current input range. TRIG:LEV 0.25 is the same as TRIG:LEV 25
PCT.
To query the analyzer's current input range, send [SENSe:]VOLT[1|2|3|4]:RANG?. To determine the
units, send VOLT[1|2|3|4]:RANG? UNIT.
The trigger source must be one of the analyzer's four input channels, TRIG:SOUR INT1, TRIG:SOUR
INT2, TRIG:SOUR INT3, or TRIG:SOUR INT4.
24-6
TRIGger:LEVel {<number>[<unit>]}|<step>|<bound>
limits:
-9.9e37:9.9e37
OUTPUT 711;"TRIGGER:LEV .10"
OUTPUT 711;"trig:lev -50 PCT"
OUTPUT 711;"TRIGGER:LEVEL 2.5 V"
TRIGger:LEVel?
Real
Option:
not applicable
Synchronization Required:
Preset State:
+0 V
SCPI Compliance:
no
confirmed
command/query