Agilent Technologies 35670A Manual page 223

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

CALCulate[1|2|3|4]:UNIT:DBReference:IMPedance
Specifies the system's reference impedance value in ohms (Ω).
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [OHM]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command specifies the system reference impedance for the dBm reference level. The dBm unit is
referenced to 1 milliwatt. Specify a value that matches the impedance of the system under test.
For example, the system impedance of a telephone system is typically 600 Ω.
The trace specifier, CALCulate [1|2|3|4], determines which trace you are selecting. Omit the specifier or
send 1 for trace A, 2 for trace B, 3 for trace C, or 4 for trace D.
CALCulate[1|2|3|4]:UNIT:DBReference:IMPedance {<number>
[<unit>]}|<step>|<bound>
limits:
1e-15:1e+15
OUTPUT 711;":calc2:unit:dbr:impedance 600"
OUTPUT 711;"CALC:UNIT:DBR:IMP 3.2e3 ohm"
CALCulate[1|2|3|4]:UNIT:DBReference:IMPedance?
Real
Option:
not applicable
Synchronization Required:
Preset State:
50 OHM
SCPI Compliance:
yes
instrument-specific
CALCulate
command/query
6-115