[SENSe:]
[SENSe:]AVERage:CONFidence
Specifies the confidence level used in equal confidence averaging in octave measurements.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [DB]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
The analyzer varies the average time constant to provide a 68% probability that the measured results are
within ± σ of the true mean value. There is a 95% probability that the results are within ± 2 σ of the true
value.
There are four values of σ available: .25 dB, .5 dB, 1 dB and 2 dB.
18-2
[SENSe:]AVERage:CONFidence {<num-
ber>[<unit>]}|<step>|<bound>
limits:
0.25:2.0
OUTPUT 711;"Average:Conf 0.25"
OUTPUT 711;"SENS:AVER:CONF 2"
[SENSe:]AVERage:CONFidence?
Real
Option:
1D1 Realtime Octave
Synchronization Required:
Preset State:
0.5 DB
SCPI Compliance:
no
instrument-specific
command/query