[SENSe:]
[SENSe:]FREQuency:RESolution:AUTO:MCHange
Specifies the maximum change permitted between the frequency response of the current measurement
point and the frequency response of the previous measurement point.
Command Syntax:
<number> ::= a real number (NRf data)
<unit> ::= [PCT]
<step> ::= UP|DOWN
<bound> ::= MAX|MIN
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command is used with auto resolution (FREQ:RES:AUTO ON).
The analyzer calculates the ratio of the frequency response of the current point to the frequency response
of the previous point. The ratio exceeds the value specified by this command, the analyzer estimates a
correction for the resolution and applies it to the next measurement point.
18-36
[SENSe:]FREQuency:RESolution:AUTO:MCHange {<number>
[<unit>]}|<step>|<bound>
limits:
0.00391:100
OUTPUT 711;"sense:freq:resolution:auto:mch 30 PCT"
OUTPUT 711;"FREQ:RES:AUTO:MCH .25"
[SENSe:]FREQuency:RESolution:AUTO:MCHange?
Real
Option:
1D2
Swept Sine
Synchronization Required:
Preset State:
+2 PCT
SCPI Compliance:
no
instrument-specific
command/query