Agilent Technologies 35670A Manual page 636

Dynamic signal analyzer gpib programming
Hide thumbs Also See for 35670A:
Table of Contents

Advertisement

TRIGger
TRIGger:EXTernal:RANGe
Selects the range for the external trigger's level.
Command Syntax:
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command works with the TRIG:EXT:LEVEL command.
When TRIG:EXT:RANG is HIGH, the allowable range is -10 V to +10 V. The external trigger level may
be set in 78.125 mV increments.
When TRIG:EXT:RANG is LOW, the allowable range is -2 V to + 2V. The external trigger level may be
set in 15.625 mV increments.
This command is only valid with external triggering (TRIG:SOUR:EXT).
24-4
TRIGger:EXTernal:RANGe HIGH|LOW
OUTPUT 711;"trig:external:rang HIGH"
OUTPUT 711;"Trig:External:Rang LOW"
TRIGger:EXTernal:RANGe?
HIGH|LOW
Option:
not applicable
Synchronization Required:
Preset State:
HIGH
SCPI Compliance:
no
instrument-specific
command/query