SYSTem:SET
Transfers an instrument state between the analyzer and an external controller.
Command Syntax:
<STATE> ::= #<byte>[<length_bytes>]<data_bytes>
<byte> ::= one byte specifying the number of
<length_bytes> ::= bytes specifying the number of
<data_bytes> ::= the bytes that define an instrument
Example Statements:
Query Syntax:
Return Format:
Attribute Summary:
Description:
This command transfers a complete instrument state—the same information contained in a state
file—between the analyzer and your controller. This allows you to store an instrument state on your
controller's file system. The state cannot be altered.
When you transfer an instrument state to the analyzer, you can use either the definite or indefinite length
block syntax. When the analyzer returns the state to a controller, it always uses the definite length block
syntax. For more information about transferring block data, see the GPIB Programmer's Guide .
SYSTem:SET <STATE>
to follow (ASCII-encoded)
to follow (ASCII-encoded)
OUTPUT 711;":SYST:SET STATE"
OUTPUT 711;"system:set state"
SYSTem:SET?
definite length <STATE>
Option:
not applicable
Synchronization Required:
Preset State:
not applicable
SCPI Compliance:
data bytes
no
confirmed
SYSTem
command/query
length bytes
state
21-25