Job Card CHK A001 - Boot/Init and Shutdown

1
SUPPLIES
None
2
TOOLS
None
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 20 minutes
4
SAFETY PRECAUTIONS.
To remove power from the complete system, switch off the external circuit breaker which con-
nects the system to the hospital supply. There are two circuit breakers inside the Generator Cab-
inet, one for the Gantry and Generator, the other for the Control Station and imaging equipment.
Switching off one of these breakers removes power from only part of the system.
Remember that the UPS will continue to supply power to parts of the Control Station and imaging
equipment until it is shut down.
NEVER switch off the UPS manually (risk of data loss).
NEVER switch off power to the Conditioner except in emergency (risk of damage to the Detector if
the correct conditions of temperature and humidity are not maintained).
WARNING
In normal operation (circuit breakers closed):
Power is present in the Generator Cabinet, MDR, and Control Station at all times, even
after the AWS, UPS, and other components have shut down. Control Station components
remain powered by the UPS for about two minutes after switch off. The 230 V supply to the
Conditioner the UPS, and the Light Box is maintained at all times.
5
PREREQUISITES
None
JC-CHK-A-001.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card CHK A001 - Boot/Init and Shutdown
Page no. 1729
Senographe DS
Chapter 9
Chapter 9

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Senographe DS and is the answer not in the manual?

Questions and answers

Dr.Robert Mahmut
March 26, 2025

G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required

1 comments:
Mr. Anderson
March 26, 2025

The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:

- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)

- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)

- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)

### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)

These values include the gantry, generator cabinet, control station, detector, and all related components.

This answer is automatically generated

Table of Contents

Save PDF