GE Healthcare
Revision 1
Job Card CHK A013 - Check mA and mAs settings
1
SUPPLIES
None
2
TOOLS
•
Radiation Protection Plate
•
Meter for measurement of mAs. The meter used must be in good condition and possess valid certi-
fication.
Suitable meters are:
-
Fluke 87 multimeter
-
Keithley 35035 mAs meter
-
GS Poner 165 mA-mAs meter
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 30 minutes
4
SAFETY PRECAUTIONS
The procedures in this Job Card produce X-rays. Ensure that you take appropriate precautions when
operating with X-rays (for example, stay behind the radiation screen during an X-ray exposure).
5
PREREQUISITES
•
The Generator covers must have been removed for access to the kV/mA board 701-PL1 on the HV
tank unit. Refer to
JC-CHK-A-013.fm
Job Card PHY A042 - Remove/Reinstall Generator Covers on page
Service Information and Procedures Class A 2385072-16-8EN
Job Card CHK A013 - Check mA and mAs settings
Page no. 1755
Senographe DS
Chapter 9
513.
Chapter 9