GE Senographe DS Manual page 1769

Hide thumbs Also See for Senographe DS:
Table of Contents

Advertisement

4
PM TASK SCHEDULE (VISITS 1-6)
Note:
Some of the tasks listed in this schedule may overlap with similar tasks required by your local regulations. If this is the case you
can skip these particular tasks and perform the tasks required by your local regulations instead.
Task Title
Inspection:
General system inspection checks
Generator/Tube:
Remove Generator Covers.
Clean Conditioner Air Filter
Check and top-up Conditioner coolant
Change Generator CPU Lithium Battery
Check HV Unit
Check Half Value Layer (HVL)
Check kVp settings
Check mA and mAs settings
Reinstall Generator covers.
Task Definition
Perform a general inspection of the entire system.
See
Job Card PM A155 - General System Inspec-
tion Checks on page
1775.
Job Card PHY A042 - Remove/Reinstall Genera-
tor Covers: p. 513
Job Card PM A429 - Conditioner Air Filter Clean-
ing: p. 1807
Job Card PM A428 - Conditioner Coolant Level
Check and Top-up: p. 1805
Job Card D/R A402 - Generator CPU Battery: p.
1549
Job Card PM A407 - Check HV Unit: p. 1801
Job Card CAL A046 - Half-Value Layer Measure-
ment: p. 1703
Job Card CHK A011 - Check kVp settings: p.
1745
Job Card CHK A013 - Check mA and mAs set-
tings: p. 1755
Job Card PHY A042 - Remove/Reinstall Genera-
tor Covers: p. 513
Int.
Drn.
PM Visit no.
mth.
min.
1
2
3
4
6
10
X
X
X
X
6
5
X
X
X
X
6
5
X
X
X
X
6
5
X
X
X
X
36
25
12
5
X
X
12
10
X
X
12
5
X
X
12
30
X
X
6
5
X
X
X
X
5
6
X
X
X
X
X
X
X
X
X
X
X
X
X
X
X