GE Healthcare
Revision 1
Job Card PM A201 - Gantry operational checks
1
SUPPLIES
None
2
TOOLS
•
Mammography compression scale with digital display
•
Set of acrylic plates
3
REQUIRED EFFORT
Personnel: 1 Field Engineer
Time: 10 minutes
4
SAFETY PRECAUTIONS
WARNING
Before beginning work on the lift mechanism, shut down the equipment and switch OFF
the Senographe hospital supply circuit breaker. Lock it in the OFF position or mark it
clearly with a warning sign to prevent others from applying power.
•
NEVER switch off at the UPS except in emergency (risk of data loss).
5
PREREQUISITES
None
6
PROCEDURE
6-1
Check lift movement
1. With the compression paddle out of compression, check that all four sets of lift up/down control but-
tons are operational (check each button for up and down movement of lift).
2. Move arm to its upper and lower limits. Software limits should stop movement. If the software limits
fail, optical sensors are used to recognize and stop any movement before impact with mechanical
stops.Total one-way movement time (going either up or down) between upper and lower limits should
be 8 ±2 seconds.
3. Lift movement should be smooth and progressive; check for any noise which might indicate exces-
sive friction or binding.
6-2
Check Arm Rotation
1. With the compression paddle out of compression, check that all four arm rotation control buttons are
operational.
For each button, check that light or firm pressure gives the correct rotation, and that rotation stops
when the pressure is released. Refer to
Keypad Functional Checks on page
2. Rotate the arm through its full range of +185°/-165°. Check in several positions that the electrical
braking is functioning correctly, by pulling on the handles.
JC-PM-A-201.fm
Service Information and Procedures Class A 2385072-16-8EN
Job Card PM A201 - Gantry operational checks
Job Card ELE A015 - Gantry Functional
580, for a full list of these checks.
Page no. 1777
Senographe DS
Chapter 11
Checks, section
6-4
Chapter 11
Need help?
Do you have a question about the Senographe DS and is the answer not in the manual?
Questions and answers
G.E.Essential l have recently bought What is the full weight.of the mamo Full weight of the console and other İtem's also measurements of each item with the main machine please All this Shipping Company required
The GE Senographe DS mammography machine is shipped in three crates with the following dimensions and weights:
- Crate 1 (Gantry, Generator Cabinet, and ramp):
- Dimensions: 2066 mm (81.34 in) depth × 848 mm (33.39 in) width × 2292 mm (90.24 in) height
- Mass: 641 kg (1410.2 lbs)
- Crate 2 (Control Station):
- Dimensions: 1110 mm (43.70 in) depth × 770 mm (30.31 in) width × 1850 mm (72.83 in) height
- Mass: 210.5 kg (464.07 lbs) for truck transport (without side covers and lid)
- Mass: 223.5 kg (492.73 lbs) for plane/boat transport (with side covers and lid)
- Crate 3 (Detector):
- Dimensions: 872 mm (34.33 in) depth × 2270 mm (89.37 in) width × 260 mm (10.24 in) height
- Mass: 104.4 kg (229.68 lbs)
### Total Mass:
- For truck transport: 641 + 210.5 + 104.4 = 955.9 kg (2104.95 lbs)
- For plane/boat transport: 641 + 223.5 + 104.4 = 968.9 kg (2133.61 lbs)
These values include the gantry, generator cabinet, control station, detector, and all related components.
This answer is automatically generated