Tools In The Fret View Options Control Block For Sensitized Emission - Zeiss LSM 880 Operating Manual

Table of Contents

Advertisement

LSM 880
Center Screen Area / Image Containers - Display and Image Analysis
6.18.2
Tools in the FRET View Options
Control
Emission
FRET tab (Fig. 700):
Within this tab overlay regions can be defined.
(The drawing tools correspond to the drawing
tools described in section Graphics).
The checkboxes Numbers and Measure refer to
the overlay regions and annotate the number of
the region and its area to the overlay in the image.
Export allows the resulting FRET image to be
saved as a separate image file.
The regions defined can be set as Object (Region where FRET should be calculated or the reference
coefficient values are determined from) or Background (from which the threshold for the image analysis
is deduced and automatically used for analysis and displayed in the Threshold tab).
Regions and the assigned status as Object or Background can be selected or deselected for the individual
analysis using the corresponding check box under Enabled.
To start analyzing the Analyze button must be clicked once. If all further changes shall be effective
immediately the On the fly checkbox can be checked.
Three different methods for FRET analysis are available in the drop down list:
Fc (Youvan),
FRETN (Gordon) and
N-FRET (Xia).
Fc or Youvan method:
Youvan et al., Biotechnology et alia 3, 1 (1997)
Fc = Ff-Df(Fd/Dd)-Af(Fa/Aa)
Displays the Fc image with intensities converted from the FRET index calculated for each pixel using the
Youvan method. This method assumes that the signal recorded in the FRET channel is the sum of real
FRET signal overlaid by donor crosstalk and acceptor signal induced by direct (donor) excitation. There is
no correction for donor and acceptor concentration levels and as a result the FRET values tend to be
higher for areas with higher intensities.
FRETN or Gordon method:
Gordon et al., Biophys J 74, 2702 (1998)
FRETN = FRET1/ Dfd * Afa
Displays the FRET image with intensities converted from the FRET index calculated for each pixel using the
Gordon method. This method calculates a corrected FRET value and divides by concentration values for
donor and acceptor. This method attempts to compensate for variances in fluorochrome concentrations
but overdoes it. As a result cells with higher molecular concentrations report lower FRET values.
10/2014 V_01
CHAPTER 1 - SYSTEM OPERATION
Block
for
Sensitized
000000-2071-464
Fig. 700
FRET View Options Control Block,
FRET tab
ZEISS
505