Zeiss LSM800-02 Quick Start Quide

Airyscan confocal microscope

Advertisement

Quick Links

Zeiss LSM800-02 + Airyscan confocal microscope
System description:
I.
Upright confocal microscope ZEN blue 2.6 HotFix-8
II.
2 GaAsP +1 AiryScan detector (GaAsP)
III.
Motorized stage for multiple position imaging
IV.
Heating chamber
V.
NO CO
mixer
2
VI.
Multi-slide holder, dish holder, single slide holder
Objectives:
- Position1: Objective Plan-Apochromat
- Position2:
Objective Plan-Apochromat
- Position3: Objective Plan-Apochromat
- Position4: Objective
- Objective Plan-Apochromat
- Objective Plan-Apochromat 40x / NA 1.2 Water (not mounted)
Laser lines: 405, 488, 561, 640 nm
Detectors:
Two very sensitive Gallium arsenide phosphide photomultiplier tube (GaAsP PMTs) detectors
with free choice of the spectral range. (Uses a tunable short pass and band pass filter in front
of the PMTs)
Be very careful with using it to prevent instrument damage! If oversaturated, detectors will
switch off automatically.
1 GaASP
PMT for the
mode (detection range adjustable by a tunable short pass and a tunable long pass filter, no
bandpass filters)
An ESID detector for transmitted light.
Simultaneous detection of 3 dyes possible.
VII. Starting up:
1. Switch on the multiple socket (1)
to the computer and microscope controllers
10x / NA 0.45 Air
20x / NA 1.0 W
20x / NA 0.8 Air
Plan-Apochromat
40x/1.4 Oil DIC
63x / NA 1.4 Oil
Airyscan
detector that can function as a normal detector in confocal
"Hanging on the right side of the
(WD=2.1mm)
DIC (WD=1.8mm)
(WD=0.13 mm)
(not mounted)
20/11/2019
rack". Provides power
1

Advertisement

loading

Summary of Contents for Zeiss LSM800-02

  • Page 1 Zeiss LSM800-02 + Airyscan confocal microscope 20/11/2019 System description: Upright confocal microscope ZEN blue 2.6 HotFix-8 2 GaAsP +1 AiryScan detector (GaAsP) III. Motorized stage for multiple position imaging Heating chamber NO CO mixer Multi-slide holder, dish holder, single slide holder...
  • Page 2 2. Switch on SYSTEM (2). 3. Switch on COMPONENTS (3). 4. Make sure that lasers are on. Laser key should be in position “I”. (4) 5. Switch on the computer. 6. Turn on Cool LED (by pressing Big blue on/off button), if you need it for visual examination and don’t forget to close the shutter when you don’t need it or at the end of your session.
  • Page 3 • Select the ‘Acquisition’ tab in the main menu of Zen blue • Go to “Smart Setup” choose your dyes (set the LUT) and configuration, you can choose between simultaneous (fastest), Best compromise line scanning mode (Smart) and sequential (best signal) scan, press ”Apply”. (Avoid Line mode as it blocks the changing of filters and sliders, therefore any change will affect all the track!)
  • Page 4 • Set pinhole to Airy Unit (AU) 1 in the channel dialog to achieve confocality. Note, that activating the ‘1AU’ button by clicking on it once (gets light blue background color) fixes the pinhole to 1AU. In this case, the match pinhole function in the Z-stack setup will not change pinhole sizes. •...
  • Page 5  Tile scan dialog will open  Set desired number of horizontal and vertical tiles  You can choose different regions to scan. Choose regions and press button “+” to add it.  Press “start experiment”  Advanced Set up allows you to do preview scan of region of interest. ...
  • Page 6 Switching off: 1. DO NOT save any “Experiment” or overwrite the default “Experiment” (marked with a home logo) as this setting will be used in the next startup of the system and might cause potential damage to GaAsP PMTs. 2. Close the ZEN Software 3.
  • Page 7: Laser Safety Instructions

    Laser Safety Instructions  During operation of class 3B and class 4 lasers red warning lights have to be switched on manually  Red warning light at the door of the room, containing laser-based equipment, prohibits the entrance  Optical path of the laser beam at all setups has to stay intact and should never be disassembled by a user.