Count Rate - Zeiss LSM 880 Operating Manual

Table of Contents

Advertisement

ZEISS
Fig. 339
Count rate window with count rate
display
Fig. 340
Overload display
Fig. 341
Count rate window with correlation
display
254
CHAPTER 1 - SYSTEM OPERATION
Left Tool Area and Hardware Control Tools
000000-2071-464
5.3.17

Count Rate

• Click the Count rate button to open the Count
rate window (Fig. 339).
a)
Display
Select one of the display modes Count rate,
Correlation
CPM
or
pressing the corresponding button. The arrow
buttons of the scale tab allow to adjust the
measuring range.
• Click the Count rate button in the Display
line to display the count rate per second (CR)
for each activated physical channel (Fig. 339).
This value is constantly updated and calculated
each 0.5 ms, if Averaging is switched Off. If
system overload occurs, an error message will
be displayed and the APDs switched off in order
to avoid damage (Fig. 340). Press Ok to
continue
and
concentration of your sample.
• Click the Correlation button to display the
correlation amplitude (G(0)) of each activated
physical and Software channel (Fig. 341). This
amplitude is constantly updated and calculated
each 0.5 ms if Averaging is switched Off. The
expanded display for each channel can be
closed by the Channel 1, Channel 2 and Cross
Correlation arrows.
LSM 880
(Counts / molecule)
laser
light
intensity
10/2014 V_01
by
or

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lsm 880 nlo

Table of Contents