The Processing tool tab (Fig. 390) is structured into
different fields.
The top field or header (just beneath the Main tool tab
area) contains the Apply button. The button is only
active when an image is selected. If the Apply button is
pressed, the selected or active processing tool operations
will be applied to that image.
The next field contains the list of Processing tools.
Select the desired tool by clicking on it. The active tool
will be highlighted.
Next is the image selection field. Highlight the
image in the image container that you want to
select. Load the Input image by pressing the
Select button. Select will load the active image in
the container. Alternatively, first press Select and
then click on the image in the image container.
With some image processing tools, two images can
be loaded. Then select the first image to be loaded
in the image container and press the Select button
for Input image. Then select the second image to
be loaded in the image container and press the
Select button for Input image 2.
Note that once an image is loaded by pressing
Select, it cannot be unloaded but only be replaced
by another one. This holds also true if the
Processing tool is changed; the previous loaded
image will stay selected. If another image is to be
loaded, just highlight that image in the image
container and press Select again. The newly
selected image will replace the old one.
If two input images are selected (Input image and
Input image 2), they have to match each other in
their dimensions. If they do not match, a warning
is given. Acknowledge the warning to continue
with the processing. The message will also be
displayed if you load a new image pair that does
not match in dimensions to the old pair. Since you
can only load one image at a time, this will lead to
a mismatch between the old Input Image 2 and
the new Input Image. Acknowledge the warning
to continue loading the second new image.
It is the responsibility of the user to check if an operation is suited for an image.
Once an image is selected by pressing the Select button, the Output Image name field with the name
of the processed image will appear. The name can be later changed when the processed image is saved.
The next field is the method specific Settings panel that can be expanded / collapsed by pressing the
triangle repeatedly. The triangle only appears when a processing tool is highlighted. It displays all the
operational options for a specific processing function, if applicable, for the loaded image. Note that some