Zeiss LSM 880 Operating Manual page 171

Table of Contents

Advertisement

LSM 880
Select the carrier configuration from the
Configurations list by highlighting it or
search for it by typing its name in the Search
field. The Info window pops up Fig. 241).
Press Yes to delete the configuration, No to
keep the configuration. To save a carrier
configuration
press
(symbolized by the disc icon).
This opens the Save current carrier
configuration as panel (Fig. 242). Type in
an appropriate name in the Configuration
field and press Ok to save the configuration
in the data base or press Cancel to abort the
process.
To assign the actual position of the stage
with a position in the sample carrier, move
to the center of a well in the actual plate
under the microscope then click Calibrate
followed by clicking the field in the sample
carrier representing this well of the plate.
This assignment will be kept until changed
repeating this procedure and that well will
turn blue.
Pressing the ? button (Fig. 243) will display a
help text how to calibrate the position of the
sample carrier.
If you click on any other than the blue
highlighted well, the stage will move to this
well based on the distances defined in the
carrier properties. This will be than the blue
highlighted one.
By pressing Select you can highlight those
wells of the sample carrier that should be
sequentially used by just clicking on them
(Fig. 244). All selected wells will turn blue. By
clicking on a selected blue highlighted well,
it will be deselected. By pressing Select all,
all wells will be selected for imaging. By
Clear
all,
pressing
deselected. As long as Select stays pressed,
the wells will be sequentially approached
starting at the let upper position to the right
lower one (A1, A2...;B1, B2,...).
10/2014 V_01
CHAPTER 1 - SYSTEM OPERATION
Left Tool Area and Hardware Control Tools
Save
the
button
all
wells
will
be
000000-2071-464
Fig. 241
Delete information window
Fig. 242
Save current carrier configuration
as window
Fig. 243
Calibration of sample carrier
Fig. 244
Selection of wells in a sample carrier
ZEISS
165