General Operation Of The Maintenance Tool - Zeiss LSM 880 Operating Manual

Table of Contents

Advertisement

LSM 880
The Maintenance Tool can only be used, if the calibration objective is entered as APO Calibration LSM.
If this is not the case, a message will remind the user to select the calibration objective.
2.3

General Operation of the Maintenance Tool

The desired task can be chosen from the list and started with the start button.
Tasks available for LSM
Sharpness Test
Scan Field Test
Scanner Calibration
Pinhole to MBS Calibration
In case a task requires any manual adjustments, the Maintenance Tool guides users to all adjustment
steps.
After the completion of a task, the result is displayed with a traffic light type green/yellow/red indicator.
• If green is indicated, the task has been fully accepted.
• If yellow is indicated, repeat the task and try to optimize the settings (e.g. check focus, KÖHLER
illumination, halogen lamps, lasers (no standby) etc.
• If red is indicated, repeat the test one more time; if it still fails, contact your local LSM service engineer.