LSM 880
After acquisition, the grey dynamics of the image are automatically linearized (see description above) to
match the range of the output display. This step visibly intensifies weak signals (Fig. 427).
Fig. 427
HDR image display using Sampling mode (left side without, right with HDR)
Unlike the Sampling method, the Illumination method of HDR still provides a linear grey stepping
which allows quantitative measurements. For the acquisition step with increased laser power, those
regions of the image are excluded from exposures when no reasonable data can be obtained with any
higher laser power.
10/2014 V_01
CHAPTER 1 - SYSTEM OPERATION
Left Tool Area and Hardware Control Tools
000000-2071-464
ZEISS
323