Ratio Channels - Zeiss LSM 880 Operating Manual

Table of Contents

Advertisement

ZEISS
Fig. 158
Channel Settings panel of a Ratio Channel
The ratio channels are displayed as additional images.
Select the required formula and enter the relevant values.
Letters can be entered into the formula fields which will be valued as 1; it is also possible to make no
entry, which will also be valued as 1, but will not be displayed.
Set by min/max allows the definition of the display scaling according to the expected minimal and
maximal values.
• Laser lines and laser power are displayed according to each track. It is possible to vary laser intensities
(in the same way as in the Imaging Setup tool).
All parameters can be varied while imaging.
120
CHAPTER 1 - SYSTEM OPERATION
Left Tool Area and Hardware Control Tools
000000-2071-464
5.2.9.2

Ratio Channels

The parameters for ratio metric imaging can
be set when a ratio channel is selected.
• Click on the button of a ratio channel
(e.g. R1).
• Clicking on the required tabs enables you
to choose from four formulas (Type 1 to
4) for ratio calculation. The relevant
decimal values can be entered in the
input boxes. The entered values remain
unchanged even after switchover to
another formula and can be reactivated
any time.
The formula type activated last is always
used for ratio formation during the scan
procedure. If the input box does not contain
any value at all or no suitable value, the
useful value last used will be activated.
LSM 880
10/2014 V_01