Loading a method
The Load button (open file symbol) will allow you to load a pre-defined method. The configurations will
be displayed; however, hardware settings change only if a measurement is triggered.
Fig. 335
Load Methods panel
• Click on the Load button.
This opens the Method panel (Fig. 335). Another click will close the window again. Scroll tabs will allow
viewing all the contents. The Methods panel lists all existing measure methods for selection or deletion.
• You can select the analysis method you want to use by highlighting the corresponding entry from the
list with a single click and pressing the Ok button.
• Press Cancel, to close the Load Method panel without loading the highlighted method.
• You can view all the meta data in the Settings display box (only available if Show all is activated). This
entry corresponds to the information of the Information on Experiment tab. Using the Copy button
you can copy the content to the clipboard. The Save button allows you to save the information in a
text file.
• All parameters will load with this method.