Acquisition Parameter - Shuttle And Find - Zeiss LSM 880 Operating Manual

Table of Contents

Advertisement

ZEISS
5.2.13
Acquisition Parameter – Shuttle and Find
Shuttle and Find allows calibration and imaging of sample positions in laser scanning microscopes (LSM)
for reproducible repositioning after transfer to scanning electron microscopes (SEM).
For Shuttle and Find different types of sample holders are pre-designed. In addition the user can define
other correlative sample holders.
The correlative sample holders have three fiducial markers enabling a three point calibration.
The calibration can be done in transmitted light or reflected light.
A motorized microscope stage is mandatory.
The workflow is unidirectional from LSM/SR to SEM only.
Fig. 174
Select Sample Carrier Template
dialog
Fig. 175
New Sample Template dialog
The distances can be determined using the Positions button in the Dimensions control view and the
functionality for Marks in the Stage tool in the Left Tool Area.
• Activate the Positions button and click on the first correlative marker.
• Determine the coordinates by clicking on Mark in the Stage tool.
• Go on with this procedure for marker two and three. Use coordinates for calculating the distances
between the markers.
132
CHAPTER 1 - SYSTEM OPERATION
Left Tool Area and Hardware Control Tools
000000-2071-464
Different types of sample holders are available
(Fig. 174), e.g.:
− Life Science cover glass 22x22
− Life Science for TEM Grids
− MAT Universal A
− etc.
The Life Science cover glass 22x22 or the
sample holder for TEM Grids should not be
used
together
Axio Examiner
adjustable
sample
calibration
procedure
neighbouring objectives will collide with the
level adjustable holder.
Creating of user-defined correlative holders
In the calibration dialog window new correlative
holder files can be defined (Fig. 175).
• Click on the + button in the Select Template
dialog.
• Type in a name for the new holder / sample
carrier. An image of the new holder can be
loaded.
In the holder file the distances (in millimeters)
between the first and the second marker and
between the second and third marker have to be
inserted.
LSM 880