LSM 880
5.3.2.1
Light Path for 2 Channel Module
(BiG.2 on DC out port)
BiG.2 has two GaAsP detectors that can be used
simultaneously to acquire two autocorrelation
functions (AC) per each channel. There are also
two cross-correlation (CC) channels available,
where the channels are correlated to each other.
Selection of dichroics and filters
From top to bottom you have the following
dichroics and filters available.
Main dichroic beam splitter (MBS):
The MBS reflects the specified laser lines and
allows the resulting fluorescence spectrum to pass
through.
• Click on the MBS icon to open a list of dichroics
(Fig. 261).
• Select the appropriate one by a mouse click.
• Close the list by clicking the icon again.
• You have two major beam splitters, one for the Visible light and one of the Invisible light (if
applicable). A plate should be placed in the dichroic that will not be used. If the system is not
equipped with a violet laser line, the Invisible light major beam splitter is not available.
Secondary dichroic beam splitter (SBS):
The SBS splits the fluorescence spectrum onto the various detection channels. The long pass (LP) filter lets
wavelengths pass above the specified border wavelength and reflects wavelengths below. The short pass
(SP) filter lets wavelengths pass below the specified border wavelength and reflects wavelengths above.
The SBS is positioned in a 45° angle to the incident beam.
The icon is inactive, as the SBS is part of the filter cube together with the emission filters, which has to be
replaced manually.
Emission filter 1 and 2 (EF1 and EF2):
The EF is used to narrow the fluorescence spectrum. The band pass (BP) filter allows the range within the
specified border wavelengths to pass through. The long pass (LP) filter allows the range above the
specified border wavelength to pass through. The short pass (KP) filter allows the range below the
specified border wavelength to pass through. The EFs are positioned in a 90° angle to the incident beam.
The secondary beam splitter within the scan head, which deflects the signal either to the 2
channel module (BiG.2) or to the 3/32 channel is set automatically to mirror/plate as needed.
10/2014 V_01
CHAPTER 1 - SYSTEM OPERATION
Left Tool Area and Hardware Control Tools
000000-2071-464
Fig. 261
Light Path tool for 2 Channel
Module
ZEISS
179