Counting (Raw Data) Format
In this display, photons are displayed according to their number (Fig. 162).
Fig. 162
Counting display in photon counting image. Settings were digital gain 1, scanning speed 8
corresponding to a pixel period or pixel dwell time 1.52 µs, 8 bit image
In the Status Area the displayed value of the pixel to which the mouse points in the image is displayed
as the number of photons.
Counting images can not be used for further processing in the LSM software like mean of ROI
function, statistics etc. They can be exported to have access to the raw data.
10/2014 V_01
CHAPTER 1 - SYSTEM OPERATION
Left Tool Area and Hardware Control Tools